University of Hertfordshire
32
To set the Start Frequency
c460-00:~ # echo MTA LISTEN 18 DATA FA 100MZ| cat -> /dev/gpib0
or shortcut as:
c460-00:~ # echoSEND 18 FA 100MZ’”| cat - > /dev/gpib0
FA is a Spectrum Analyser command used to set the Start Frequency. In this case, the Start
Frequency is set to 100 MHz.
To set the Stop Frequency
c460-00:~ # echo MTA LISTEN 18 DATA FB 300MZ| cat -> /dev/gpib0
or shortcut as:
c460-00:~ # echoSEND 18 FB 300MZ| cat - > /dev/gpib0
FB is a Spectrum Analyser command used to set the Stop Frequency. In this case, the Stop
Frequency is set to 300 MHz.
To set the Zero Span mode
c460-00:~ # echo MTA LISTEN 18 DATA SP 0HZ| cat -> /dev/gpib0
or shortcut as:
c460-00:~ # echoSEND 18 SP 0HZ| cat - > /dev/gpib0
SP is a Spectrum Analyser command used to Changes the total displayed frequency range
symmetrically about the center frequency. Specifying 0 Hz enables zero-span mode, which
configures the spectrum analyzer as a fixed-tuned receiver.
To get the Spectrum Analyser model number
c460-00:~ # echo MTA LISTEN 18 DATA ID?| cat -> /dev/gpib0
or shortcut as:
c460-00:~ # echoSEND 18 ID?’”| cat - > /dev/gpib0
Przeglądanie stron 38