Step 8: Auto-calibration is finished (Fig. D-8)
Figure D-8: D/A Calibration is finished
D/A channel Manual-Calibration
Step 1: Click the Manual D/A Calibration tab to show the D/A channel
manual calibration panel. Eight D/A channels are individually
calibrated . Before calibrating, output desired voltage from the
D/A channels and measure it through an external precision
multi-meter.
Step 2: For example, choose channel 0; select the Range and select the
wished output voltage code or value from the radio buttons (Fig.
D-9).
Figure D-9: Selecting D/A Range